GE Healthcare · CTLightSpeed2450476010002300110068350001139560721789815
LightSpeed 5.X (80/100kW)
Symptom
"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors
Error message
Error: 245047 Unable to allocate scan files See the long description for the full error code Thu May 12 17:43:37 2005 Error: 245047 Host: rrmcct16 Process: scanmgr File: scan_sequence.cxx Line: 1461 Function: System Control : Scan Control Scan Type: Scout Scan: 0/0/0 Scan Seq Id: 601000 Unable to allocate scan files Error: 230011006 Unable to create directory on scan data disks. Requested scan can not be allocated. See long description for complete error Fri Feb 10 08:38:41 2006 Error: 230011006 Host: mct2 Process: scanmgr File: sdio_xfs_io_ops.c Line: 230 Function: Data Acquisition : SBC Processing Scan Type: Scout Scan: 0/0/0 Scan Seq Id: 835000 Unable to create directory on scan data disks. Requested scan can not be allocated. Scan: MCT2/1834/1/1 Directory name: /raw_data/ex1834.MCT2.1139560721.789815 Permissions: 775 (octal) Unix system error : 140 --- Remote I/O error. Error: 245215 Scan setup failed on Step:7 (Get Cal Module). Status:-1 See Long Description for complete error Tue Oct 11 10:45:52 2005 Error: 245215 Host: er_ct5_vct Process: scanmgr File: scanmgr_lib.cxx Line: 1224 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 524000 Scan s
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2362873 SDDA (Scan data disk array) used on GOC3. See Long Description for details. The GOC3 to GOC4 upgrade kits are now under 1 part number, 5144058-500 plus the FE will have to also order the DARC2 that is needed by the DAS as a separate order. Part #: 2362873 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SCAN DATA DISK ARRAY - GRE FRU: Yes Status: Orderable Comments: OBSOLETE PART case 8391074 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 05-23-07 ** FE ASSIST ** For CT not PET.. IF no stock of 2362873 SDDA must upgrade to DARC II + ICOM Kit per one of the following part number's * 5144058-500, 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL * 5144058-501, 4,8 MDAS DARC Replacement Kit, JARRELL * 5144058-502, 4 SDAS DARC Replacement Kit, JARRELL Documentaion 5194645-100 Must order software release 06MW29.7, p/n 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better (2/19/07 212001261) *** 2ND OPINION NOTES *** Look at the DARC messages log for SCSI disk errors located at /var/log/messages
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.