Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2600010235072326000235850011260003117260007391

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Do Not Use - use one of the other two Error: 260001023 See Long Description for complete error NOTE: The error message code included in this error is variable. You will need to refer to the log to find the cause of this error. Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 07:23:22:910 Log Series: 37 The DCB has notified the system of an error condition which prevents scanning. See log for further information: Error message code: 260002358 Notification level: 5 Sequence number: 4 Fri May 11 06:44:04 2007 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 374 Function: No System Function Reported Scan: 50011/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:44:05:145 Log Series: 42 The CCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260003117 Notification level: 5 Sequence number: 4

Error: 260007391 The Data Acquisition System Has Detected An Error In The -/+12V analog DAS Power Supply See LD for more details Tue Aug 8 07:49:0

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues