GE Healthcare · CTLightSpeedVCT26010720650180260107204
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260107206 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. See Long Description for full error Wed Apr 7 21:04:02 2004 Error: 260107206 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3771 Wed Apr 7 21:03:47 2004 Host : Table Gantry Processor Ermes # : 260107206 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3771 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. (Cmd: 27 Current State:0 Flags:0x8 ReInit Status:3). or Host : Table Gantry Processor Ermes # : 260107206 Exception Class : Abort Severity : Secondary File : amc mgr.c Line# : 3883 Function : No System Function Reported Scan Type : Axial Scan: 50180/1/2 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. (Cmd: 21 Current State:0 Flags:0x8 ReInit Status:3) Error: 260107204 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry. See Long Descript
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5125429 FUSE KIT for VCT systems See Long Description for details 5125429 FUSE KIT for VCT systems NOTICE: The information below came from a document that is under revision control and it is possibly out of date. Always check for the latest revision of the manual if necessary . # NUMBER QTY FRU DESCRIPTION/NAME 1 2142415-4 3 Yes PROTECTION FUSE, 5A, TIME DELAY, 600VAC 250VDC, BOARDS USE FUSE CLIPS 2 2142415-5 3 Yes PROTECTION FUSE, 8A, TIME DELAY, 600VAC 250VDC, BOARDS USE FUSE CLIPS 3 2142415-7 3 Yes PROTECTION FUSE, 15A, TIME DELAY, 600VAC 250VDC, BOARDS USE FUSE CLIPS 4 2142415-9 2 Yes PROTECTION FUSE, 25A, TIME DELAY, 600VAC 250VDC, CHASSIS MOUNTED-- CURRENT RATING HIGHER THAN FOR TYPICAL WITH FUSE CLIPS 5 2336517-2 2 Yes FUSE 25 AMPS 700 VOLTS 2.5M SECONDS 6 2351492 3 Yes 200A FUSE 7 2364059 2 Yes GLASS FUSE 8 2379651 2 Yes FUSE - 700 V, 200 A, FAST SEMI 9 46-170021P106 2 Yes FUSE 8 AMPS 250 VOLTS BUSSMANN ONLY 10 46-170021P75 2 Yes 015.000A 0250V 3AB SLOW UL BUSSMANN MDA 15 11 46-267217P20 4 Yes FUSE METRIC SLO BLO 2.500 A 250V 5X20MM GLASS LITTLEFUSE 218 02.5 12 46-267217P23 2 yes FUSE METRIC SLO BLO 5.000 A 250V 5X20MM GLASS LITTLEFUSE 218 005 13 5118644 1 Yes FAST ACTION 15 AMPS 250 VOLTS 0.01 SECONDS U0028FF 14 U8006FF 3 Yes FUSE 1.0A 15 U8008FF 3 Yes FUSE 2.0A 16 U8009FF 3 Yes FUSE 3.2A 17 2390269 3 Yes FUSE 2.5 AMPS 125 VOLTS
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.