GE Healthcare · CTLightSpeed200002352
LightSpeed 5.X (60kW)
Symptom
System will not boot to or stay in applications. System will not boot or stay in applications
Error message
Error: 200002352 HOST SHUTDOWN was initiatied by the unexpected termination of scanfilemgr. Error: 200002352 Host: sjhodact1 Process: cupMonitor File: ssProcStop.c Line: 766 A HOST SHUTDOWN was initiatied by the unexpected termination of scanfilemgr.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2362870-10 Computer - Data Acquisition Recon Control, 16-slice MDAS and GDAS, 2U, EMC Ed2 Compliant. See Long Description Part #: 2362870-10 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 2.18 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer - Data Acquisition Recon Control, 16-slice MDAS and GDAS, 2U, EMC Ed2 Compliant FRU: Yes Status: Orderable Comments: ** FE Assist ** applies to CT Not Pet systems.If Your System has SP2 loaded must remove prior to replacement Refer to Service Note SN09052911a .(6/5/09 21201261) If no stock is available order a Kit 5144058-500 : 16 MDAS 16,8,4 GDAS DARC REPLACEMENT KIT, JARRELL Refer to Dir 5194645-100, Must have 06MW29.7 or LATER. For Pet Westville DARC2 with P/N 5114772 and ICOM assembly 2404320. (2/05/2009 12001261) added per EMC2 eco 2007993 as replacement for 2362870, ms13452 041015
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.