Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26013319350179260133169260133189

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260133193 - Filter A failed to Home (or Filter B failed to Home) See long description for complete error code Sun Mar 18 05:49:17 2007 Error: 260133193 Host: orp Process: orp_subsys File: filter_home.cxx Line: 158 Sun Mar 18 05:49:16 2007 Host : Collimator Ermes # : 260133193 Exception Class : Abort Severity : Primary File : filter_home.cxx Line# : 158 Function : No System Function Reported Scan Type : None/Unknown Scan: 50179/7/0 Filter A failed to Home. OR it could be filter A and B. Sun Mar 18 05:49:17 2007 Error: 260133193 Host: orp Process: orp_subsys File: filter_home.cxx Line: 165 Sun Mar 18 05:49:16 2007 Host : Collimator Ermes # : 260133193 Exception Class : Abort Severity : Secondary File : filter_home.cxx Line# : 165 Function : No System Function Reported Scan Type : None/Unknown Scan: 50179/7/0 Filter B failed to Home.

Error: 260133169 drive monitoring detected the current motor position is out of the tolerance band for a successfull move. See long description for complete error Mon Jan 29 13:54:09 2007 Error: 260133169 Host: orp Process: orp_subsys File: ccb_drive_timers.cxx Line: 599 Mon Jan 29 13:54:08 2007 Host : Collimator Ermes # : 260133169 Exception Clas

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues