Pular para o conteúdo principal
GE Healthcare · CTHiSpeed186218

CT/e Dual, CT ProSpeed AII/FII Irix or Linux

Symptom

Scanner unable to generate X-RAY Exposure. Scanner unable to generate X-RAY Exposure

Error message

Error: 186218 40-0104H Rotation current open load Rotation board detected that no current is flowing to the motor (rotor). XG JEDI error code 40-0104H Rotation current open load Rotation board detected that no current is flowing to the motor (rotor). Fri May 7 12:29:20 2010 Error: 186218 Host: FMH0 Process: tgp_out File: tgp_in_input.c Line: 334 XG Error phase is 3 code is 40. detail error code : 40 - 0104H Rotation error : Rotation Openload

Applies to

Models
HiSpeed
Product
CT/e Dual, CT ProSpeed AII/FII Irix or Linux HiSpeed CT/e Dual CT Scanner System see long Description for sales information Product Name: CT/e Dual CT Scanner System Irix A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) N/A Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5392132 Programmed Rotation Board - replaces 2409043, 2242440-11, 2214605-3 in Jedi Generator Auxiliary Module (This rotation brd. used on ALL HPower Systems and YMS Hispeed) See Long Description for Details. 5392132 Programmed Rotation Board - replaces 2409043, 2242440-11, 2214605-3 in Jedi Generator Auxiliary Module (This rotation brd. used on ALL HPower Systems and YMS Hispeed) Part #: 5392132 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRU- Programmed Rotation V3 board FRU: Yes Status: Orderable Comments: ** FE Assist ** For P500D or Xray RAD, order part number 2409043. CT and Vascular continue to order 5392132. 212010438 3/1/11 17Aug2010 - Promoted for ECO 2104218 212006824 for 305024364. Part #: 2409043 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRU-PROGRAMMED ROTATION BOARD FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues