Symptom

System will not boot to or stay in applications. System will not boot or stay in applications

Error message

Error: 200002352 A HOST_SHUTDOWN was initiated by the death of procICE. See the Long Description for the full error code Fri Oct 7 08:29:28 2005 Error: 200002352 Host: ct03 Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the death of procICE.

Error: 200002372 The process procICE took too long to start up. See long description for full error Tue Mar 30 15:07:53 2004 Error: 200002372 ct01 cupMonitor ssProcStart.c 839 The process procICE took too long to start up.

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2197234-2 RIP / Motorola 2300 Series Board for ICE Box See long description 2197234-2 RIP / Motorola 2300 Series Board for ICE Box
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.