GE Healthcare · CTLightSpeed260134802398728100000
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry
Symptom
Cradle / IMS on GT1700 / GT2000 would not advance from home and makes grinding noise. See Long Description for more details and sound. Cradle / IMS on GT1700 / GT2000 would not advance from home and makes grinding noise This problem happen after moving a large patient off of the table, or after physically moving the IMS to it’s mechanical out of gantry limit (home). Attachment is the sound the table makes when the cradle move button is pushed. The button was pressed 2 times (2 sounds)
Error message
Error: 260134802 IMS position is out of tolerance. See Long Description error messages Thu Oct 20 08:01:13 2005 Error: 260134802 Host: tgp Process: tgp subsys File: gtcb movechk.cxx Line: 581 Thu Oct 20 08:01:13 2005 Host : Table Sub-System Ermes # : 260134802 Exception Class : Secondary Severity : Pri/Soft File : gtcb movechk.cxx Line# : 581 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 IMS position is out of tolerance. current position(um) : 398728 target position (um) : 0 start position (um) : 398728 velocity (um/s): 100000 time from start (ms) : 300
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5168472-2 New style IMS bearing replacement kit for GT1700 or GT2000 tables. See long description for details Part #: 5168472-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: IMS DRIVE REPLACEMENT SET Positioning GT FRU: Yes Status: Orderable There is a new Kit for the VCT Table IMS. The old kit is no longer available. The old Kit part number will automatically change up to the number for the new kit: 5168472-2 . FE's are confused by the parts in the Kit and there are no installation instructions. FMI 26809 will be released with instructions: The Kit contains only one Bearing, but the Table has 2 Bearings. This is correct . The new Bearing is larger and as a result only one Bearing is required now. There are no instructions in the Kit on how to install all the parts. Use the hyperlink below to get instructions on how to install the Kit. http://jptyoiis01.ap.health.ge.com:8000/ctdoc/FMI/19/26809.pdf
- Area
- CT