Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2001090122001070232001091002005200152000004164294967295

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Unable to display any images; exam browser listing in image works is empty and fails to refresh.

Error message

Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed. Cannot build scan request.

Error: 2052 error during execution of the fonction (note: may or may not be misspelled) See LD for details Tue Jun 7 12:27:18 2005 Error: 2052 hostid: su101 process: imserver imsprim.c 595 error during execution of the fonction /usr/g/sdc_image_pool2/reserved Input/output error Note: fonction is not mis-spelled Wed Sep 12 10:26:07 2007 Error: 2052 Host: gect Process: imserver File: imsprim.c Line: 622 error during execution of the function /usr/g/sdc_image_pool/reserved Input/output error

Error: 200107023 A Failure had been detected while calling UIR x_ToScanRequest function or srx_sm_generate_action function See Long Description for error variations Tue Oct 19 09:22:20 2004 Error: 200107023 Host: dgct Process: scanRx File: srx_sma_manage_scan.c Line: 688 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling UIRx_ToSca

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required