GE Healthcare · CTLightSpeed2450476010002452155240002910000
LightSpeed 5.X (60kW)
Symptom
"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors
Error message
Error: 245047 Unable to allocate scan files See the long description for the full error code Thu May 12 17:43:37 2005 Error: 245047 Host: rrmcct16 Process: scanmgr File: scan_sequence.cxx Line: 1461 Function: System Control : Scan Control Scan Type: Scout Scan: 0/0/0 Scan Seq Id: 601000 Unable to allocate scan files Error: 245215 Scan setup failed on Step:7 (Get Cal Module). Status:-1 See Long Description for complete error Tue Oct 11 10:45:52 2005 Error: 245215 Host: er_ct5_vct Process: scanmgr File: scanmgr_lib.cxx Line: 1224 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 524000 Scan setup failed on Step:7 (Get Cal Module). Status:-1 Fri Apr 27 12:05:20 2007 Error: 245215 Host: jrct Process: scanmgr File: scanmgr_lib.cxx Line: 1154 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 2910000 Scan setup failed on Step:3 (Allocate Scan File). Status:245047
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS See LD for more info 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS Part #: 5114772 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DARC Computer - Dual Disk H16 MDAS, GDAS FRU: Yes Comments: Added per GOC4 RSPL Replace the DARC2: 1) If you get a Westville DARC P/N 5114772 load the DARC software to recover. 2) If you get Jarrell p/n 5147442 perform LFC using 06MW29.7 software.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required