Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Ring artifact on axial images with no errors in logs - document in RFS - critical issue See long description for artifact Ring artifact on axial images with no errors in logs

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2336885 MDAS 4 or MDAS 8 Convertor Board, replaces 2322994-3. See Long Description for details. 2336885 MDAS 4 or MDAS 8 Convertor Card, replaces 2322994-3 (see below) Part #: 2336885 Part Type: GP CONSUMABLES Part Weight: 0.48 LBR Repairable: No Returnable: No Hazardous Material: N Description: SERVICE, MDAS CONVERTOR BD FRU: Yes Status: Orderable 4 Slice versus 8 Slice Configuration The HiSpeed QX/i system can operate 4 slice per rotation, data collections mode, whereas, the Detector and MDAS are capable of operating 4 slice or 8 slice per rotation. From a hardware perspective the main difference is in the MDAS. The 4 slice configuration uses a depopulated MDAS. This means that all EVEN slots are empty. Only ODD numbered slots contain converter boards. Data collection is still performed in the same manner using only the odd converter card slot assignments.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues