GE Healthcare · Nuclear MedicineInfinia17693917001701484
Infinia Hawkeye X-Ray Diagnostic, CT monitor 6 failed System Diagnostics -> X-Ray -> X-Ray Detector -> Full DA…
Infinia Hawkeye X-Ray Diagnostic, CT monitor 6 failed
Symptom
Infinia Hawkeye X-Ray Diagnostic, CT monitor 6 failed System Diagnostics -> X-Ray -> X-Ray Detector -> Full DAS Graph Figure 1 - Full DAS Graph Failure Figure 2 - CT monitor 6 failed
Error message
CT_DIAG_APPLY_CORRECTIONS_ERROR (176939) CT-Error trying to apply corrections to raw data CT_DIAG_APPLY_CORRECTIONS_ERROR (176939) -- CT-Error trying to apply corrections to raw data - Extract from LOG file Syslog.csv Fatal error: CT_DIAG_APPLY_CORRECTIONS_ERROR:.:CT-Error trying to apply corrections to raw data:.:1) Try again. 2) Perform Full shutdown and try again.:.: CT-Error trying to apply corrections to raw data:.:Trace:Caused by: [email protected] 15:41:02 ERROR = Error in Activity ID:0 Stack trace:com.ge.med.fi.tacqUtil.diag.manager.DiagnosticException: Detector test failed on : monitors test (DIAG_DETECTOR_FAILED,170017,ERROR) at com.ge.med.fi.tacqUtil.diag.manager.quasar.ct.DASFunctionalDiag.runDiagnostic(DASFunctionalDiag.java:217) at com.ge.med.fi.tacqUtil.diag.manager.FruDiagnostic$DiagRunnerThread.run(FruDiagnostic.java:275) Source:not stated Description:com.ge.med.fi.tacqUtil.diag.manager.DiagnosticException: Detector test failed on : monitors test (DIAG_DETECTOR_FAILED,170017,ERROR) [email protected] 15:40:53.421 [th:01484] ERROR=Error in CTReconServer d:\tacq\su\src\cpp\ct\hawkeyems\image\recon\corrections\offset\offset.cpp(29): offset::Have bad monitor c
Applies to
- Models
- Infinia
- Product
- Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
- Component
- *X-Ray (Generator, DAS, Controller) *X-Ray (Generator, DAS, Controller)
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.