Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000745650201

ID: 141166 LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Fastcals failed to complete. Fastcals failed to complete

Error message

Error: 260007456 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance. See LD for more details Error: 260007456 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance Mon Mar 20 16:36:22 2006 Error: 260007456 Host: DAS Process: DCB Firmware File: zaxis.c Line: 578 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50201/5/19 Type: Axial Exception Level: Pri/Soft Time: 16:36:22:380 Log Series: 83 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance. Tolerance : 200 umd Control Error : 319 umd Beam Side : A Side

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5142849-4 SCAN WINDOW LIGHTSPEED SERIES H2 THRU HP RIM COVERS (replaces 2101967-4). See long description for details. Part #: 5142849-4 Part Type: GP CONSUMABLES Part Weight: 7.94 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASSEMBLY, SCAN WINDOW, LIGHTSPEED SERIES H2 THRU HP RIM COVERS, WEATHERSTRIP UPDATED MAY 2005 FRU: Yes Status: Orderable Comments: This part is replacement for 2101967-4, please discontinue use of 2101967-4 and use only 5142849-4. Added per ECR2014906, ms13452 050713
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required