Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260112016 Gantry forward touch sensor was pressed , Gantry can only tilt in the forward direction till XXX deg See Long Description for complete error. Mon Jan 21 14:56:59 2008 Error: 260112016 Host : Table Gantry Processor Ermes # : 260112016 Exception Class : Secondary Severity : Pri/Soft Gantry forward touch sensor was pressed at -12500 mdeg. Gantry can only tilt in the forward direction till -10000 mdeg. Error: 260112014 Touch sensor ? Touch Sensor was PRESSED See Long Description for full error Error: 260112014 Touch sensor ? Touch Sensor was Pressed. Sun Mar 12 15:14:25 2006 Error: 260112014 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr_gt.c Line: 9013 Sun Mar 12 15:14:24 2006 Host : Table Gantry Processor Ermes # : 260112014 Exception Class : Secondary Severity : Sec/Soft File : uif_table_tilt_mgr_gt.c Line# : 9013 Function : PATIENT ANATOMY POSITIONING : Tilt Drive Scan Type : None/Unknown Scan: 0/0/0 Touch sensor BACK Touch Sensor was PRESSED. Elevation position where touch Sensor was Pressed:0 um. Tilt position where touch sensor was pressed:0 mdeg. Remote Tilt button is DISABLED. Use local gantry controls to tilt the gantry. Or Mon Apr 9 09:29:44 2007 Er
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.