Symptom

Table elevation numbers fluctuate up and down on their own or sometimes may not change at all during elevation Table elevation numbers fluctuate up and down on their own or sometimes may not change at all during elevation.

Error message

Error: 260134703 Elevation moved without command. See LD for more details Wed Jun 14 12:48:34 2006 Error: 260134703 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 530 Wed Jun 14 12:48:34 2006 Host : Table Sub-System Ermes # : 260134703 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 530 Function : No System Function Reported Scan Type : None/Unknown Scan: 412 Elevation moved without command. If characterize completed appears above, this isnt error. If this appears many times, check potentiometer. Current position = 903212 um Last stop position = 897741 um Tolerance = 5000 um Time since last stop = 751380 msec

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5127774 ELEVATION POTENTIOMETER ASSY-VCT See Long Description for details Part #: 5127774 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION POTENTIOMETER ASSY FRU: Yes Status: Orderable Comments: source pole/planner code revised per to Global Table transfer
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.