Pular para o conteúdo principal
GE Healthcare · CTLightSpeed183190

LightSpeed 4.X - Octane

Symptom

System would not scan with tube ID parameter in Rotor Characterization file does not agree with tube. System would not scan with tube ID parameter in Rotor Characterization file does not agree with tube .

Error message

Error: 183190 Tube ID parameter in Rotor Characterization file does not agree with tube.Tube ID parameter in Rotor Characterization file = 8. Actual tube ID =0. Rotor Characterization file rejected. See the Long Description for the complete error code Tue Oct 11 01:19:16 2005 Error: 183190 Host: OBC Process: Generator File: gen_in_mail.c Line: 1521 Function: XRAY GENERATION : Tube Identifiers Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 01:18:54:420 Log Series: 2 9 Tube ID parameter in Rotor Characterization file does not agree with t ube. Tube ID parameter in Rotor Characterization file = 8. Actual tube ID = 10. Rotor Characterization file rejected. Or Error: 183190 Host:OBC Process: Generator File:gen_in_mail.c Line: 1521 Function: XRAY GENERATION : Tube Identifiers Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 23:15:36:265 Log Series: 27 Tube ID parameter in Rotor Characterization file does not agree with tube. Tube ID parameter in Rotor Characterization file = 8. Actual tube ID =0. Rotor Characterization file rejected.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2179860-2 HEMRC CONTROL BOARD See LD for more info 2179860-2 HEMRC CONTROL BOARD. Part #: 2179860-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HEMRC CONTROL BOARD FRU: Yes Status: Orderable FEMC PART COMMENTS: FLS - HEMRC CONTROL BOARD. REPLACES P/N 2179860. ADDED PER PCN 220717.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.