Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26011836733328118796926011823654034

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118367 - An unexpected JEDI state change occured - JEDI not ready See long description for full error Thu Jun 12 07:08:56 2008 Error: 260118367 Host: orp Process: orp_subsys File: gen_exec.c Line: 863 Thu Jun 12 07:08:55 2008 Host : On-Board Rotating Processor Ermes # : 260118367 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 863 Function : No System Function Reported Scan Type : Scout Scan: 33328/1/1 An unexpected JEDI state change occured . JEDI not ready -- Tcks: 1187969 Line: 7420

Error: 260118236 - (70-0501) DC bus out of range Wed Nov 26 12:56:16 2003 Error: 260118236 Host: orp Process: orp_subsys File: gen_exec.c Line: 1923 Wed Nov 26 12:56:14 2003 Host : Jedi Ermes # : 260118236 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 1923 Function : No System Function Reported Scan Type : Static Scan: 54034/1/4 (70-0501)DC bus out of range. Z6,Z7= 0h,0h

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
2336517-2 Fuse 25 A 700 V 2.5M seconds - used in Jedi Generator See Long Discription for Details Part #: 2336517-2 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 25 AMPS 700 VOLTS 2.5M SECONDS FRU: Yes Status: Orderable Comments: This part is used in Fuse Kit's for Hpower (2218570-2), Hp60 (2385412)and VCT 5125429). Please check order as Fuse Kits, ms13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required