Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT185122182902

LightSpeed 2.X (Plus, QX/i) - GRE

Symptom

Table will not move. Table will not move

Error message

Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Power is OFF

Error: 182902 Emergency Off button has been pressed Sun Sep 5 16:04:39 2004 Error: 182902 Suite: RP21 Host: ETC Process: Table Controls File: tc emergency.c 1.6 2/16/99 Method: No Method line: 89 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 4016 Log Series: 22 Emergency Off button has been pressed

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2236720 TABLE TOP COVER ASSEMBLY - LEFT See long description for details Part #: 2236720 Part Type: GP CONSUMABLES Part Weight: 7.26 KGM Repairable: No Returnable: No Hazardous Material: N Description: COVER ASSEMBLY TABLE TOP - LEFT FRU: Yes Status: Orderable Comments: FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FLS - CAN ALSO ORDER OLD COVER 2149735-2 GPC TABLE MOVE TO CHINA 5/24/0NOW ASIA SOURCED. PER PCN 220864, THIS PART HAS BEEN REPLACED BY P/N 2372371, IN ORDER TO BE USED ON HPOWER SYSTEMS. ALSO SOURCED BY ASIA. MS205444 3/31/03. DO NOT DISCOUNT - PER SHAKUN RIGHT: 2236720-2 ONE USE LIGHTSPEED
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues