GE Healthcare · CTLightSpeed244596
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Gantry reset failed. Version Mismatch was Detected between a Firmware Controller and the SRU Disk. Gantry reset failed. Version Mismatch was Detected between a Firmware Controller and the SRU Disk
Error message
Error: 244596 A Version Mismatch was Detected between a Firmware Controller and the SRU Disk See long description for full error Error: 244596 Host: _sbc Process: fwmgr File: HwSetUp.c Line: 986 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 A Version Mismatch was Detected between a Firmware Controller and the SRU Disk. Firmware File: /usr/g/fw/alms/OBCRScan.bin
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 303.1_H3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 303.1_H3M4 303.1_H3M4 for LightSpeed 3.x (H2 Gantry with MDAS) Obsoleted by FMI 25313 which in turn was Obsoleted by FMI 25331 303.1_H3M4 for LightSpeed 3.x (H2 Gantry with MDAS) Obsoleted by FMI 25313 which in turn was Obsoleted by FMI 25331 Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2305092-5 303 Ver 303H3.0M4 2305093-5 303 Ver 303 H3.0M4 2305091-5 2305090-5 Ver 303H3.0 M4 Apps ver 303.1_H3M4.1 FW08-02 Obsoleted by FMI 25313
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required