GE Healthcare · CTHiSpeed20010901220050200120000182116558080551936
HiSpeed QX/i - Octane
Symptom
Scanner Hardware Stopped Scan with Image Allocation errors. Scanner Hardware Stopped Scan with Image Allocation errors
Error message
Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed. Cannot build scan request. Error: 200502001 Failed to send Image Space Allocation Request with Status = 200001821. See long description for full error Error: 200502001 Failed to send Image Space Allocation Request with Status = 200001821. Total space of 16558080 (bytes), image size is 551936 (bytes).
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-78UU or 2169940-78 Octane/Octane II Computer 18 GB Disk Drive. See long description Part #: 2169940-78UU Part Type: GP USED Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCT2 18GB 10K RPM DRIVE FRU: Not Available Status: Orderable Stan Pearson, Asset Mgmt move to used material program 6-27-05 Part #: 2169940-78 Part Type: GP CONSUMABLES Part Weight: 1.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCT2 18GB 10K RPM DRIVE FRU: Yes Status: Orderable Comments ** FE ASSIST ** 2169940-12 is no longer available can use 2169940-72 9GB or a 2169940-78 18GB Hard drives for a CTi or lightspeed QXI Octane systems as Image or OS disk. Will need to use updated LFC per 2282382-100 rev 2 or 2284291-100 rev 8. The 2169940-78 has 5000 images files where normal is 20000 if used as a CTi Image Disk. Refer to Service Note 0711261b. (212001261 8/20/08)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.