Symptom

Scanner hardware reset and flash download fails

Error message

Error: 182827 TNC Start Input to ETC FAILED. Possible Communication Problem. Gantry Push Button, Foot Pedals, and Remote Tilt Might NOT Function. Mon May 2 15:51:45 2005 Error: 182827 Host: ETC Process: Gantry Display File: gd_gantryCtrls.c Line: 452 Running Version Management Tool Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:51:45:790 Log Series: 42 TNC Start Input to ETC FAILED. Possible Communication Problem. Gantry Push Button, Foot Pedals, and Remote Tilt Might NOT Function.

Error: 182825 HW/SW error: Communication Problems with the GANTRY DISPLAY. Open Long Description for full error Wed Dec 22 20:49:26 2004 Error: 182825 Host: ETC Process: Gantry Display File: gd_tnc_driver.c Line: 471 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:48:40:485 Log Series: 30 HW/SW error: Communication Problems with the GANTRY DISPLAY. Number of Communication Timeout errors: 0. Number of NACK's errors: 3. Number of DUART errors: 0. Numver of Checksum errors: 0.

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2256730-3 Display assembly used on 2.X/3X/4X5X/7X See long description for details. Part #: 2256730-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: H-POWER COMPATIBLE GANTRY DISPLAY FRU: Yes Status: Orderable Comments: FLS - H-POWER COMPATIBLE GANTRY U/O CT H POWER DISPLAY
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.