Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2100004072005200152000004164294967295200000415

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Unable to display any images; exam browser listing in image works is empty and fails to refresh.

Error message

Error: 210000407 Image space query failed. See Long Description for more details Mon Feb 27 07:19:00 2006 Error: 210000407 Host: csbipct2 Process: dms File: Initialize.cxx Line: 1085 Image space query failed. Error code: -1

Error: 3015 Send message failed :QRS.9301 See long description for full error Tue Nov 27 10:38:09 2007 Error: 3015 Host: tediivct Process: netserver File: netproc.c Line: 357 Send message failed :QRS.9301

Error: 200520015 Updating image space, received a bad status code: See the long description for different status codes Sat Jan 22 18:23:06 2005 Error: 200520015 Host: chr09bc1cti Process: environmentMgr File: ImageSpaceFormMgr.c Line: 117 Updating image space, received a bad status code: 200000416 Reason: Response from SdC timeout you may also see this error with different status codes, such as Fri Jul 28 05:12:48 2006 Error: 200520015 Host: vct1 Process: environmentMgr File: ImageSpaceFormMgr.c Line: 121 Updating image space, received a bad status code: -1 Reason: OS error : -1 --- Unknown error 4294967295 Wed Nov 16 08:23:54 2005 Error: 200520015 Host: gect_qxi Process: environmentMgr File: ImageSpaceFormMgr.c Line: 121 Updating image space, received a bad 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required