Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2000027182639023000650733928200002748230009041

LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

Image reconstruction failures. Image reconstruction failures

Error message

Error: 200002718 Image create Pixel data request failed See the long description for the full error code Thu Jun 30 23:03:01 2005 Error: 200002718 Host: ip01 Process: imagecreate File: ic imagecreate req.c Line: 247 Image : 26390/2/3 Recon Sequence Id : 15 Image create Pixel data request failed

Error: 230006507 Reconmgr timed out waiting for IG done from IG Control See long description for a full sample error. Tue Aug 19 12:10:21 2008 Error: 230006507 Host: ice Process: reconmgr File: rm process job.c Line: 817 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 33928/2/1 Image: 33928/2/84 Recon Seq Id: 991 Reconmgr timed out waiting for IG done from IG Control SDC Setup:102 SDC Done:0. IG Setup:200 IG Done:0.

Error: 200002748 Timeout waiting for pixel data response from icproxy See long description for a full sample error. Tue Aug 19 12:10:33 2008 Error: 200002748 Host: mrct2mid Process: imagecreate File: ic getpixeldata.c Line: 324 Image : 33928/2/83 Recon Sequence Id : 990 Timeout waiting for pixel data response from icproxy

Error: 230009041 The image did not complete reconstruction within the time allotted. See long description for complete error 

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console - showprods returns 308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console - showprods returns 308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console
Area
CT

Full cause & step-by-step resolution