GE Healthcare · CTLightSpeedVCT260132912260132914
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Customer can not scan with DAS temperature errors. Customer can not scan with DAS temperature errors
Error message
Error: 260132912 DAS Digital Module Temperature Out of Tolerance See Long Description for more details Wed Mar 22 00:23:01 2006 Error: 260132912 Host: orp Process: orp subsys File: ThermalManager.cxx Line: 2282 Wed Mar 22 00:23:00 2006 Host : Das Ermes # : 260132912 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 2282 Function : No System Function Reported Scan Type : None/Unknown Scan: 7/2/0 DAS Digital Module Temperature Out of Tolerance Digital Module: 28 Temperature: 5003 Limits: 3800, 5000 Error: 260132914 Rail Temperature Out Of Range See Long Description for complete error Temperatures vary. Wed Jun 21 14:51:53 2006 Error: 260132914 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 2071 Wed Jun 21 14:51:52 2006 Host : Das Ermes # : 260132914 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 2071 Function : No System Function Reported Scan Type : None/Unknown Scan: 81/201/0 Rail Temperature Out Of Range Sensor: Rail Low Channel Temperature: 3640 <--- In this case, temp was always too LOW Average A/D Temperature: 0 Limits: 3650, 3950
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5272157-2 - Cherokee 48Vdc 1KW supply with Ferrite filter - used on 7.X Gantry Stationary frame - replaces 5125186 and 5272157 See Long Description for details Part #: 5272157-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Cherokee 48Vdc 1KW supply with Ferrite filter FRU: Yes Status: Orderable Comments: Per ECR 2086962, 212008898, 7/9/09. Part #: 5272157 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Cherokee 48Vdc 1KW supply with modified input filtering. FRU: Yes Status: Orderable Comments: Per ECR 204282, 212008898, 12/4/07 Part #: 5125186 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PM3326A-6-1-2-C Pioneer Magnetics 48V 1kW Power Supply FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041209 48 volts DC power supply used on VCT Stationary frame. If Rotating frame 48 volts DC supply is needed, order 5126259. (added by 212001261 9/12/05)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required