Pular para o conteúdo principal
GE Healthcare · CTHiSpeed26011822650000

HiSpeed Dual, ProSpeed FII Series

Symptom

Scan Aborts with JEDI XG Error. Scout Scan Aborts with JEDI XG Error

Error message

Error: 260118226 (60-0504 Inverter Gate Power Supply error. gate supply voltage has dropped below the level required to drive the IGBTs properly. see Long Description for full error Wed Oct 11 16:27:45 2006 Error: 260118226 orp orp_subsys gen_exec.c 2376 Wed Oct 11 16:27:43 2006 Host : Jedi Ermes # : 260118226 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Static Scan: 50000/1/1 (60-0504)Inverter Gate Power Supply error. gate supply voltage has dropped below the level required to drive the IGBTs properly. Z6,Z7= 0h,0h

Applies to

Models
HiSpeed
Product
HiSpeed Dual, ProSpeed FII Series HiSpeed Dual, ProSpeed FII Series Product Name: Hispeed Dual ( Linux ) CT/Be2 A list of types. (a Product/hardware table) This console has an HP8000, HP8200, or HP8400 Computer (not GRE console) An On-site / Remote identification aid . (Example: check_config) HP Host computer A Table of options (included, optional, new introductions, …) HiSpeed Dual for China and Global B71402CA has the following as standard Options 1.5 sec fast scan, Power 200ma, Thichness 1mm, Raw data 5000, Image 100000 A list of Software that has run on this product (Product / Software Compatibility Table) With HP8000/8200 : • OS Version 4.5.15144980•Software Version 2.235165193 (except China)•Software Version 2.235172510 (for China only)•Patch 2.245179238•Patch 2.255197828•Patch 2.275267772 With HP8400 : •OS Version 6.2.95265475•Software Version 3.105270457•Patch 3.305267336•DVD Interchange 6.15270187•Patch 3.40 (See Note 8)5316222•DVD Interchange 7.05316345N/ Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5126989 Jedi Xi Series Inverter Mid Power assembly use ONLY with glass tube (replaces 2215814) See long description for details Part #: 5126989 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: INVERTER MID POWER ASSY_FRU FRU: Yes Status: Orderable Comments: Use only with for GLASS X-ray tubes
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required