LightSpeed 4.X - GRE

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183118 - FIL/MA fault detected: Filament current less than 1 Amp. See long description for full error Thu Oct 28 12:14:31 2004 Error: 183118 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 337 Function: XRAY GENERATION : Filament Selection Scan: 21676/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 12:14:32:370 Log Series: 7 FIL/MA fault detected: Filament current less than 1 Amp. Desired focal spot = 1. (0 = large, 1 = small) Fault latch address = ffdf43 hex, Bit = D4

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296701P1 UNREGULATED FILAMENT POWER SUPPLY See long description Part #: 46-296701P1 Part Type: GP CONSUMABLES Part Weight: 4.55 LBR Repairable: No Returnable: No Hazardous Material: N Description: UNREGULATED FILAMENT POWER SUPPLY FRU: Yes Status: Orderable Comments: **GPRS DISCOUNTABLE PART** 212018867 19th June 09
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.