GE Healthcare · CTLightSpeedVCT
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Popup box comes up reading as follows: "importimage: cannot reserve space. Check available disk space and error logs." See Long Description for details System still works but runs slower than normal. Customer sees a message p opup box coming up continually which reads as follows "importimage: cannot reserve space. Check available disk space and error logs." System still works but runs slower than normal.
Error message
Error: 9038 tuple not found : Ims_open See LD for complete Error Fri Jan 14 07:28:36 2005 Error: 9038 Host: ct01 Process: importimage File: importimage.c Line: 519 tuple not found : Ims_open Error: 2106 Image Storage Full See Long Description for complete error Wed Aug 30 16:07:53 2006 Error: 2106 Host: IRIS Process: imserver File: imsprim.c Line: 742 Image Storage Full
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.