Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCTBrightSpeed260118159502622601182235007426011822410105

LightSpeed 5.X (80/100kW)

Symptom

Scan aborts intermittently with mA out of range. Scan aborts intermittently with mA out of range

Error message

Error: 260118159 (27-0907)Filament aging saturated. Z6,Z7= 0h,0h See LD for complete error Wed Mar 8 04:38:44 2006 Error: 260118159 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Wed Mar 8 04:38:43 2006 Host : Jedi Ermes # : 260118159 Exception Class : Soft Severity : Pri/Soft File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Axial Scan: 50262/5/48 (27-0907)Filament aging saturated. Z6,Z7= 0h,0h

Error: 260118223 (60-0402)mA scale error See Long Description for complete error Error 260118223 orp orp_subsys gen_exec.c 2213 Host : Jedi Ermes # : 260118223 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2213 Function : No System Function Reported Scan Type : Static Scan: 50074/1/9 (60-0402) mA scale error. mA has been measured to be either too low or too high with respect to mA demand 20 ms after the beginning of the exposure. Z6,Z7= 0h,0h

Error: 260118224 60-0403 mA out of range See Long Description for complete error Thu Jun 9 14:36:46 2005 260118224 orp orp_subsys gen_exec.c 2150 Host : Jedi Exception Class : Soft Severity : Pri/Soft File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Hel

Applies to

Models
LightSpeed, VCT, BrightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.