Pular para o conteúdo principal
GE Healthcare · CTHiSpeed183179

Scan abort during any scan with external x-ray interlocks not satisfied.

Symptom

Scan abort during any scan with external x-ray interlocks not satisfied.

Error message

Error: 183179 I/O board sensed external x-ray interlocks not satisfied. Sun Sep 5 11:10:28 2004 Error: 183179 Host: OBC Process: Generator File: gen_control.c 1.39 Line: 658 Function: XRAY GENERATION : Xray Exposure Interlocks Scan: 4167/1/1 Type: Scout Exception Level: Pri/Soft Time: 11:10:29:165 Log Series: 41 I/O board sensed external x-ray interlocks not satisfied. Possible causes: - Scanning room door is open (if applicable). - Check relay on DIP and/or Axial board. This error is flagged by the OBC on I/O board during xray generation.

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2302509-2 SIGNAL BRUSH BLOCK FOR HSA/CTI SYSTEMS See long description for details Part #: 2302509-2 Part Type: GP CONSUMABLES Part Weight: 0.62 LBR Repairable: No Returnable: No Hazardous Material: N Description: SIGNAL BRUSHBLOCK - REPLACEMENT CT/I FRU: Yes Status: Orderable Comments: FLS - 8-7-03, CONSUMABLE COLLECTION PER ANJANA ON PROMOTION THROUGH JUNE 30, 2004 DO NOT CHANGE PRICE PER SHAWN (MAYND) DO NOT DISCOUNT - PER SHAKUN Note: 2302509 FOR H1 (LIGHTSPEED) SYSTEMS That is, DO NOT mix and match ETC and Schliefring Signal Brush Blocks. Any matched pair of Signal Brush Blocks may be installed with either Power Brush Block. See HSA/CTI Non EMC and EMC Gantry Slipping Comparison in Direction 46-018316 for any updates to this information.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues