Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Error: 244584 Collimator/DAS reset failed See Long Description for complete error Error: 244584 Host: qx-ct Process: fwmgr File: HwManager.c Line: 1536 Collimator/DAS reset failed.

Do Not Use - use one of the other Error: 260001008 The OBC failed to connect with the ? (DCB and/or the CCB) sub-system in the allotted time Open Long Description to see error example Mon Mar 28 03:39:48 2005 Error: 260001008 Host: OBC Process: HOP File: hop_node_init.c Line: 120 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 03:39:46:170 Log Series: 31 The OBC failed to connect with the DCB (could be the CCB) sub-system in the allotted time. Check to see if subsystem is alive. Timeout value: 40000mS

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution