GE Healthcare · CTHiSpeed260118204
CT/e Dual, CT ProSpeed AII/FII Irix or Linux
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118204 (60-0306) No kV feedback on anode side from - CT TIC See LD for error explanation Code 60 Exposure Errors Error code: 60-0306H Message/explanation: No kV feedback on anode side kv measured <12kv 0,5ms after start of exposure on anode side only Potential cause: HV cable short circuit HV tank to PPC kV control board flat cable HV tank PPC kV control board Recommended action/ Troubleshooting guide: Remove anode HV cable on generator side and verify that there is no short-circuit on cable between pins and ground. Verify flat cable connection between PPC kV control and HV tank. Remove anode and cathode cables on generator side and run no load HV diagnostic (fill receptacle with oil). If test fails, replace PPC kV Control board or HV tank.
Applies to
- Models
- HiSpeed
- Product
- CT/e Dual, CT ProSpeed AII/FII Irix or Linux HiSpeed CT/e Dual CT Scanner System see long Description for sales information Product Name: CT/e Dual CT Scanner System Irix A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) N/A Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2201184-2 H.V. Anode Cable for use on YMS systems between HEMIT Tank and Tube See long description for details Part #: 2201184-2 Part Type: GP CONSUMABLES Part Weight: 4.23 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9230TE HV CABLE FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required