GE Healthcare · CT186416186431
CT/e, CT Prospeed AI/FI Irix or Linux
Symptom
Intermittent Scan Hardware fault
Error message
Error: 186416 TGP(MP) did not receive any reply from TGP(GP) in response to GNTRY CMD. (TGP fault code: 11-10F0-04) See Long Description for Details. Fri Jul 21 18:10:20 2006 Error: 186416 Host: NXI Process: tgp_out File: tgp_errors.c Line: 494 TGP Task Time Out. Code : 11-10F0-04 Error detected by : TGP(MP) TGP(MP) did not receive any reply from TGP(GP) in response to GNTRY CMD. Error: 186431 OGP Encoder pulse input from Aperture is out of specification while setting Aperture width. See long description for full error Sat Sep 15 11:30:28 2007 Error: 186431 Host: LINUX Process: tgp out File: tgp errors.c Line: 324 Aperture Error. Code : 20-1020-03 Error detected by : OGP Encoder pulse input from Aperture is out of specification while setting Aperture width.
Applies to
- Product
- CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
- Component
- 2257217 P2004LW APERTURE ASSY - used on CTe See long description for details Part #: 2257217 Part Type: GP CONSUMABLES Part Weight: 7.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: P2004LW APERTURE ASSY FRU: Yes Status: Orderable Comments: FLS - APERTURE ASSY USED ON CT/E AND YMS HISPEED LXI,FXI,DXI,ZXI MS08992 4/2/00 P/N 2257217 IS BACKWARD COMPATIBLE WITH 2244124. 2244124 IS NOT FORWARD COMPATIBLE WITH 2257217. FMI 22700 UPDATED THE APERTURES WITH P/N 2257217. MS26561 DSM 11/2/00
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required