Symptom

Archiving function for the MOD is grayed out and disabled from use. Archive button on the browser does not respond to cursor clicks.

Error message

Error: 10213 Semaphore creation failed Invalid argument See long description for an example error message. Thu Oct 11 10:33:12 2007 Error: 10213 Host: ct01 Process: arserver File: arsserver.c Line: 806 Semaphore creation failed Invalid argument

Error: 9047 arserver Initialization failed See the long description for the full error code Mon Oct 17 07:48:17 2005 Error: 9047 Host: nsct Process: arserver File: server.c Line: 326 Initialization failed

Error: 10062 Using DICOM as preferred save format. See long description for an example error message. Thu Oct 11 10:33:01 2007 Error: 10062 Host: ct01 Process: arserver File: arsserver.c Line: 1125 Using DICOM as preferred save format.

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5167410 SCSI Tower with MOD and DVD drives for GOC series consoles (replaces 2368838 ) See Long Description for details Part #: 5167410 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Peripheral Tower, dual-bay, MOD-DVDram FRU: Yes Status: Orderable Comments: ** FE ASSIST ** If ordering 5167410 (new) SCSI tower to replace 2368838 (old) SCSI tower for issues not related to the MOD, contact FMI Admin (Sahu, Shankar, sso 501375480 ) to order FMI 25408 (kit# 5317459) instead. FMI 25408 re-uses MOD, so if that is the failure, then continue with 5167410 order.(212001261 11/10/08) *** Second Opinion *** Reconfig should be tried prior to declaring DOA/FOA part (212001261 7/10/09) Per ECR 2030398, 8/2/06, MS11892.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.