Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000681723391260006815

LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

Scanning aborts with axial encoder errors

Error message

Error: 260006817 Axial Motor Control (AMC) Logic detected Gantry (Axial) MOTION without Command Current Position: 108.7 degs Commanded Position: 90.0 degs Degrees Tol: +/- 1 See the long description for the full error code Mon Dec 6 09:27:53 2004 Error: 260006817 Host: STC Process: Axial File: amc_mon.c Line: 987 Function: AXIAL CONTROL : axial control Scan: 23391/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 09:27:53:795 Log Series: 829 Axial Motor Control (AMC) Logic detected Gantry (Axial) MOTION without Command Current Position: 108.7 degs Commanded Position: 90.0 degs Degrees Tol: +/- 1 Retry Rx to verify probable Gantry Balance or Brake or Encoder Hardware. problem.

Error: 260006815 Axial Motor Control (AMC) Logic could did NOT position the Gantry within tolerance. See long description for full error Mon Dec 6 09:27:53 2004 Error: 260006815 Host: STC Process: Axial File: amc_exec.c Line: 1494 Function: AXIAL CONTROL : axial control Scan: 23391/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 09:27:54:170 Log Series: 830 Axial Motor Control (AMC) Logic could did NOT position the Gantry within tolerance. Current Position: 108.7 degs Commanded Position: 90

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues