Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230019002245047601000

LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708

Error: 245047 Unable to allocate scan files See the long description for the full error code Thu May 12 17:43:37 2005 Error: 245047 Host: rrmcct16 Process: scanmgr File: scan_sequence.cxx Line: 1461 Function: System Control : Scan Control Scan Type: Scout Scan: 0/0/0 Scan Seq Id: 601000 Unable to allocate scan files

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-10 73 GB destroked to 18GB 15000 Used on Octane OC with 9Gb OS hard drive. See Long Description for more detail. This part is used on Octane Operators Consoles that have a 9Gb OS hard drive and need a destroked Scan Data Disk. See SN0711061b for jumper settings. Part #: 2225004-10 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB destroked to 18GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues