GE Healthcare · CTLightSpeedVCT1387353918843531272082716802114
LightSpeed 4.X - GRE
Symptom
Images fail to display and recon shutdown. Images fail to display and recon shutdown
Error message
SCSI error : return code = 0x8000002, Deferred sdb: sense key Hardware Error, Buffer I/O error on device sdb11 See LD for more of the Errors OC: /var/log/messages Jul 6 09:02:10 gectsp kernel: SCSI Error: (0:0:0) Status=02h (CHECK CONDITION) Jul 6 09:02:10 gectsp kernel: Key=4h (HARDWARE ERROR); FRU=02h Jul 6 09:02:10 gectsp kernel: ASC/ASCQ=19h/00h "" Jul 6 09:02:10 gectsp kernel: CDB: 2A 00 08 44 EF 1F 00 00 30 00 Jul 6 09:02:10 gectsp kernel: SCSI error : <2 0 0 0> return code = 0x8000002 Jul 6 09:02:10 gectsp kernel: Deferred sdb: sense key Hardware Error Jul 6 09:02:10 gectsp kernel: Additional sense: Defect list error Jul 6 09:02:10 gectsp kernel: end_request: I/O error, dev sdb, sector 138735391 Jul 6 09:02:10 gectsp kernel: Buffer I/O error on device sdb11, logical block 8843531 Jul 6 09:02:10 gectsp kernel: lost page write due to I/O error on sdb11 Jul 6 09:02:39 gectsp kernel: SCSI Error: (0:0:0) Status=02h (CHECK CONDITION) Jul 6 09:02:39 gectsp kernel: Key=3h (MEDIUM ERROR); FRU=00h Jul 6 09:02:46 gectsp kernel: ASC/ASCQ=0Ch/02h "" Jul 6 09:02:46 gectsp kernel: CDB: 2A 00 08 44 EF 27 00 00 28 00 Jul 6 09:02:46 gectsp kernel: *********************************************
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required