GE Healthcare · CTHiSpeed97210650012450591056001245065
HiSpeed CT/i 4.X with Indigo and HSA Gantry
Symptom
System will perform a scout but fails to perform an axial or helical scan. System will perform a scout but fails to perform an axial or helical scan
Error message
Error: 97210 Database error encountered See Long Description for complete error. Error: 97210 Suite: CT01 Host: SBC0 Process: hdrSrv File: ScnAlloc.m 1.24 Method: installScn: Line: 1759 Database error encountered for Exam 65001, Series 11, Scan 34. idbErr: (1044) Record key is defined in DB dbErr: (1510) Unable to install image Error: 245059 Error Encountered Communicating with Header Services Error: 245059 Suite: CT01 Host: SBC0 Process: scanmgr File: do_axial_complete.m 1.16 Method: do_axial_complete() Line: 387 Function: Data Acquisition : SBC Processing Scan: 65001/11/4 Type: Xray-on Axial Scan Seq: 1056001 Error Encountered Communicating with Header Services. Error: 245065 Aborting xrays during axial because scanComplete message to hdrSrv failed Error: 245065 Suite: CT01 Host: SBC0 Process: scanmgr File: do_axial_complete.m 1.16 Method: do_axial_complete() Line: 393 Function: Data Acquisition : SBC Processing Scan: 65001/11/4 Type: Xray-on Axial Scan Seq: 1056001 Aborting xrays during axial because scanComplete message to hdrSrv failed
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 4.4.19.1 for HiSpeed CT/i with with CT/I (RHAPSODE) CONSOLE (Indigo) - showprods returns Rhapsode Product 4.4.19.1 4.4.19.1 for HiSpeed CT/i with with CT/I (RHAPSODE) CONSOLE (Indigo) - showprods returns Rhapsode Product 4.4.19.1 Intro. Date 10/18/00 Mandatory FMI 25268 Rhapsode Product 4.4.19.1 software for HiSpeed CT/i with SGI Indigo computer Intro. Date 10/18/00 Mandatory FMI 25268 Application Software Set Set Part No. . 2283723-3 Apps. CD Part No 2282440-3 Apps. Ver. 4.4.19M4 Apps. CD Label HiSpeed CT/I App SW showprods/Service Desktop Rhapsode Product 4.4.19.1 Class C CD Part No. Ver. CD Label 2282441-2 4.419M4 HiSpeed CT/I Class C Software Class M CD Part No. Ver. CD Label 2282442-2 4.419M4 HiSpeed CT/I Class M Software Patch Info: P/N 2318818, 4.5 Patch Rhapsode Product 4.4.19.1 Rhapsode patch 4.5.3 ver 1 11/21/2001 Optional FMI 25287A Also available via AutoSC ,_,Number:,2283723-3_______, ,BOM,Rev:,___,PI Rev:,___,CR Nb/Typ:,_______,___, ,CTZ,US, ,JAROSCH,ANN,K , ,Rev:,000,PI Rev:,000,CR Nb/Typ:, , , ,Description/Name:,E,HISPEED CT/I OPER SFWR VER 4.4.19 M4, ,Date:,2000-11-13, ,AIL:, , ,R,S,S,F, ,0001,/,0002, ,Mark, ,Number, ,Quantity,,U/M,U,T,P,C,Description/Name, ,__,1 ,2282439-2 , 1.000,PC ,S,N,N,N,IRIX 6.2M4 OS FOR 4.4 SOF, ,__,2 ,2282440-3 , 1.000,PC ,S,N,1,1,HISPEED CT/I APP SFWR VER, , , , , , , , , , , ,
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.