Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000777426000793750444

LightSpeed 5.X (80/100kW)

Symptom

Scan abort during alpha calibration. Scan abort during alpha calibration

Error message

Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on this error. Sometimes the actual bad board will be the one that is not listed.

Error: 260007937 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Long Description for full error. Tue Dec 28 09:41:32 2004 Error: 260007937 Host: orp Process: orp_subsys File: brdflt.c Line: 82 Tue Dec 28 09:41:31 2004 Host : Das Ermes # : 260007937 Exception Class : Soft Severity : Sec/Soft File : brdflt.c Line# : 82 Function : No System Function Reported Scan Type : None/Unknown Scan: 50444/1/0 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Following Error Message for Conve

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2318808-3 150 PIN - 5 connector INTERPOSER Replaces part 2318808-2 used in MDAS-16 and GDAS-16 (Quantity = 5) 2318808-3 150 PIN - 5 connector INTERPOSER Replaces part 2318808-2 used in MDAS-16 and GDAS-16 systems. Part #: 2318808-3 Part Type: GP CONSUMABLES Part Weight: 1.09 LBR Repairable: No Returnable: No Hazardous Material: N Description: (5) CONNECTORS, 150 PIN INTERPOSER FRU: Yes Status: Orderable Comments: FLS - CONNECTORS (5). ADDED PER CT PCN 220814. REPLACES P/N 2318808-2 SINCE THIS -2 PART CHANGE FROM FRU YES TO NO. MS20544 5/20/03 PCN NOTES: 2318808-3 IS A TUBE THAT CONTAINS 5 PIECES. FOR QUANTITY 1 ORDERED FE WILL RECEIVE 5 INTERPOSERS. VENDOR INFO MILL-MAX MFG. CORP S596-00-000-00-955000 VC 462905 8536.69.4040 HTS CODE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.