GE Healthcare · CTLightSpeed2601120112601151851272744134927175000260115236
LightSpeed 5.X (60kW)
Symptom
Grinding noise coming from the table while driving cradle in or out. Table may move slower than normal. Customer may hear a grinding noise coming from the table while driving cradle in or out. Table may move slower than normal.
Error message
Error: 260112011 Received Table Rx but the Table was NOT READY Error : 260112011 Received Table Rx but the Table was NOT READY Thu Sep 1 08:49:43 2005 Error: 260112011 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 3197 Thu Sep 1 08:49:44 2005 Host : Table Gantry Processor Ermes # : 260112011 Exception Class : Soft Severity : Pri/Soft File : uif_table_tilt_mgr.c Line# : 3197 Function : No System Function Reported Scan Type : Helical Scan: 8439/4/1 Received Table Rx but the Table was NOT READY for Scanning. Error: 260115185 Cradle velocity is out of tolerance Error. 260115185 Cradle velocity is out of tolerance cradle_position(um) : 1272744 target_position(um) : 1349271 velocity (um/s) : 75000 fpga status : 0x8104 Error: 260115236 MTCB Cradle Drive did not stop when expected Tue May 30 16:57:09 2006 Error: 260115236 Host: tgp Process: tgp_subsys File: mtcb_cradle_monitor.c Line: 487 Tue May 30 16:57:08 2006 Host : Table Sub-System Ermes # : 260115236 Exception Class : Soft Severity : Pri/Soft File : mtcb_cradle_monitor.c Line# : 487 Function : No System Function Reported Scan Type : None/Unknown Scan: 17521/3/0 MTCB Cradle Drive did not stop when expected. Target p
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required