GE Healthcare · CTLightSpeed200002632
LightSpeed 5.X (80/100kW)
Symptom
Date on System changed to some day in the year 2001, or earlier, after shutdown and reboot.
Error message
Error: 200002632 System Back dated. SW Options will be disabled. See Long Description for Details. Mon Dec 31 18:03:15 2001 Error: 200002632 Host: darc Process: dataacq File: swokutl.cxx Line: 784 *** Error: System Back dated. SW Options will be disabled. Please change the system date to current date or contact OLC
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2334572 The 3 VOLT BATTERY 1" DIAMETER for CPU's aka CR2032 See long description for details 2334572 The 3 Volt Battery, 1" in diameter, for CPU's also known as CR2032 This may be purchased locally at a variety of locations, including Radio Shack, Walgreens, etc. Part #: 2334572 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 3 VOLT BATTERY 1" DIAMETER FRU: Yes Status: Orderable Comments: FLS - NEWARK # 13C1019 # MANUF. CPI CANADA VC#726749 PART# 742290100 3 VOLT BATTERY 1" DIAMETER U/O LORAD M4 MICROPROCESSOR MS29328 - LMG - 2/18/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required