GE Healthcare · CTBrightSpeed26000311250284
BrightSpeed HPower or Delight HPower
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 260003112 CAM A Move Is Out of The Retry Tolerance. See long description for full error Wed Nov 5 04:16:21 2008 Error: 260003112 Host: orp Process: orp_subsys File: ccb_drive_move.c Line: 370 Wed Nov 5 04:16:19 2008 Host : Collimator Ermes # : 260003112 Exception Class : Secondary Severity : Sec/Soft File : ccb_drive_move.c Line# : 370 Function : No System Function Reported Scan Type : Axial Scan: 50284/1/1 CAM A Move Is Out of The Retry Tolerance. Actual Position : 177um -805cnt Expected Position: 46um -837cnt Retry Tolerance : 74um
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5422592 (replaces 2125453-2) Coupling and Motor Assembly, RoHS Helios Collimator - used on LS 2X - 5X collimators see long description for details Part #: 5422592 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Coupling and Motor Assembly, RoHS Helios Collimator FRU: Yes Status: Orderable Old FRU Compatible No Comments Added per to ECR2118136 by sso#305024364 Aug 30, 2011 Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands No Use Demand History Yes Ultimate Prime 5422592 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Direct ECO Number ECR2118136 Part #: 2125453-2 Description: HELIOS COLLIMATOR, CAM MOTOR ONLY FRU: Yes Status: Orderable These motors are used in collimators for LS 1X - 5X
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.