GE Healthcare · CTLightSpeedBrightSpeedHiSpeed101911689610180
LightSpeed 4.X - GRE
Symptom
MOD Save or Restore functions fail. MOD Save or Restore functions fail Troubleshooting There is no media in the drive message appears in a attention box during a MOD Save or Restore of with Error: 10191 Failed save exam. posted in the Message Log. The fllowing pop-up message appears on screen. Problem is consistent If a different disk is used then the error occurs again when trying to save, restore, or label
Error message
Error: 10191 Failed save exam. Mon Nov 27 03:45:02 2006 Error: 10191 Host: emc1 Process: arserver File: arscommand.c Line: 2304 Failed save exam. Paused 16896 Error: 10180 Resetting ARS SONYSMO-F551 See LD for the full error Error:10180 Resetting ARS SONYSMO Host: scan Process: arsjob File: ArchiveDevice.c Line: 4615 Resetting ARS SONYSMO-F551 - $Id: @(#)ArchiveDevice.oN32 vobadm
Applies to
- Models
- LightSpeed, BrightSpeed, HiSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2318598 Sony MOD Drive type SMO-551-SD which allows writing to 1.2GB media See Long description for details Part #: 2318598 Part Type: GP REPAIRABLES Part Weight: -82269.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: SONY MOD SMO-F551-SD FRU: Yes Status: Orderable Comments: OBSOLETE PART case 9240897 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 05-23-07 RDC determined that 80 percent of MOD repairs are due to dirt and dust. Extend the life of your MOD by doing the following: 1) Verify enclosure that MOD is to be installed in is cleaned before replacement. 2) Verify customer media not contaminated with dust. 3) Damaged or dirty customer media will shorten MOD life and should not be reused. Carefully slide open media window and inspect. 4) Avoid leaving media in MOD for long time periods since dust and dirt migrate into the MOD and media; MOD life shortened. 4) If media manually removed from MOD using hole in front bezel then cycle power to reset mechanism BEFORE attempting to insert next media. Instances where archiving stops and message in error log indicates Failed Saved Exam. Paused xxxx. (where xxxx is exam number) can oftentimes be fixed through software and usually does not indicate bad MOD. See Error 10191 Failed Save Exam. Paused in Problem/Solution DataBase (PSDB) for fix details. CDT 1/08/2007.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.