Release (reset) the emergency push button
Symptom
Table will not move
Error message
Alarm 099 See long description Code 099 refers to the absence of the command of the power transformer connection relay. All movements are disabled, the angle of tilt display flashes and shows the alarm code while the remaining displays switch off. Can be caused by any switch in the safety circuit being open, this includes the 2 EStop switches, II touchguard switches, and table tilt and elevate limit switches SID optical sensor switch.
Applies to
- Models
- Precision
- Product
- Precision RXi
- Area
- X-ray R&F