Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300000675000416800015000

LightSpeed 5.X (60kW)

Symptom

"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors

Error message

Error: 230000067 Unexpected event occurred during scan data transfer See the long description for the full error code Sat Apr 17 20:34:13 2004 Error: 230000067 Host: darc Process: dataacq File: dip_thread.cxx Line: 390 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Xray-on Static Scan: 50004/4/1 Scan Seq Id: 168000 Unexpected event occurred during scan data transfer. Event bit: 7 Count: 1 Description: Global bus error: read attempted from the Test Data FIFO Scan aborted. OR Tue Mar 6 07:13:50 2007 Error: 230000067 Host: darc Process: dataacq File: vdip.cxx Line: 817 Function: Data Acquisition : Scan Data Acquisition Scan Type: Cine Scan: 50004/6/1 Scan Seq Id: 15000 Unexpected event occurred during scan data transfer. Event bit: 22 Count: 1 Description: DMA interrupt was not cleared Scan aborted.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2333611 4.X-5.X 850 MB RF RECEIVER See Long Description Part #: 2333611 Part Type: GP CONSUMABLES Part Weight: 0.14 LBR Repairable: No Returnable: No Hazardous Material: N Description: 850 MB RECEIVER FRU: Yes Status: Orderable Comments: FLS - 850MB RECEIVER ADDED PER RSPL. USED ON H16. MSA8429 06/26/02
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.