Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260125034

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 260125034 Did not reboot TCB within 20sec. See Long Description for complete error Tue Aug 8 21:50:05 2006 Error: 260125034 Host: tgp Process: tgp_subsys File: ta_func_gt.c Line: 735 Tue Aug 8 21:50:04 2006 Host : Table Gantry Processor Ermes # : 260125034 Exception Class : Secondary Severity : Pri/Soft File : ta_func_gt.c Line# : 735 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Did not reboot TCB within 20sec.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 See long description for details 5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 Part #: 5126615 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASTEC Digital PS Assembly FRU: Yes Status: Orderable Comments: Added per RT / GDAS RSPL
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues