GE Healthcare · CTLightSpeed2601163075100151000
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
Hardware reset fails with Laser indicator flashing on the Gantry Display board. Hardware reset fails with Laser indicator flashing on the Gantry Display board
Error message
Error: 260116307 Could not send message to 101 or 201 from 101 or 201 (any combination) see long description for error details. Error: 260116307 Could not send message to 101 or 201 from 101 or 201. See examples below: Fri Aug 15 15:01:01 2008 Error: 260116307 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 513 Fri Aug 15 15:00:48 2008 Host : Table Gantry Processor Ermes # : 260116307 Exception Class : Secondary Severity : Sec/Soft File : lan_watchdog.c Line# : 513 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Could not send message to 201 from 201. ( TGP from TGP) Event Attempting to send 51001 System Error - 0 Sun Jul 5 13:49:14 2009 Error: 260116307 Host: tgp Process: tgp subsys File: lan watchdog.c Line: 568 Sun Jul 5 14:00:47 2009 Host : Table Gantry Processor Ermes # : 260116307 Exception Class : Secondary Severity : Sec/Soft File : lan watchdog.c Line# : 568 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Could not send message to 101 from 201. (ORP from TGP) Event Attempting to send 51000 Tue Dec 15 14:28:25 2009 Error: 260116307 Host: orp Process: orp subsys File: lan watchdog.c Line: 568 Tue Dec 15 14:24:2
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218570-2 H-POWER FUSE KIT used on Lightspeed 5x systems. See Long Description for contents of fuse kit. Part #: 2218570-2 Part Type: GP CONSUMABLES Part Weight: 43.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER FUSE KIT FRU: Yes Status: Orderable Comments: Not Available Fuse KIT contents Mark Number Name Qty Description 1 2351492 2.0 200A FUSE 2 2364059 2.0 GLASS FUSE 3 46-170021P50 2.0 "FUSE 12 AMPS 250 VOLTS BUSSMAN MDA12 DUAL ELEMENT." 4 46-170021P106 2.0 FUSE 8 AMPS 250 VOLTS BUSSMANN ONLY 5 2336517-2 2.0 FUSE 25 AMPS 700 VOLTS 2.5M SECONDS 6 46-170021P52 3.0 "FUSE 3 AMPS 250 VOLTS(REVIEWED TK, 7/91)." 7 46-170021P10 2.0 "006.000A 0250V 3AG FAST UL/C 8 46-170021P101 1.0 "FUSE 20 AMPS 700 VOLTS 2~ LONG X 9/16~ DIA FAST BLOW" 9 46-170021P11 2.0 "7.0A, 250V SLO-BLO FUSE. *TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 10 46-170021P31 2.0 "1/2A, 250V SLO-BLO FUSE. TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 11 2106993-5 2.0 FUSE 20 AMPS 500 VOLTS .3 SECONDS 12 46-327160P1 2.0 "12.0A, 125VAC, DUAL ELEMENT TIME DLY FNM-12, 0.406 X 1.5 LG, FIBER BODY" 13 2379651 1.0 FUSE - 700 V, 200 A, FAST SEMI
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required