Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT245226108400026010013724457251004

LightSpeed 5.X (60kW)

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 245226 Scanmgr requested hardware reset. Scanning hardware might be down, or the communication with the firmware might have failed. See LD for complete error Sun Sep 10 05:29:21 2006 Error: 245226 Host: cpro Process: scanmgr File: scanmgr_lib.cxx Line: 1450 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1084000 Scanmgr requested hardware reset. Scanning hardware might be down, or the communication with the firmware might have failed.

Error: 260100137 Host caused the firmware subsystems to SHUTDOWN. See long description for full error Mon May 26 07:00:32 2008 Error: 260100137 Host: tgp Process: tgp_subsys File: scHostInMail.c Line: 438 Wed Dec 31 19:00:03 1969 Host : Table Gantry Processor Ermes # : 260100137 Exception Class : Secondary Severity : Sec/Soft File : scHostInMail.c Line# : 438 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Host caused the firmware subsystems to SHUTDOWN.

Error: 244572 UnExpected Event received sender:orp_subsys event:51004 time:0 See long description for full sample error message. Mon Aug 11 08:36:32 2008 Error: 244572 Host: h0150ct1 Process: fwmgr File: fwmgr_events.c Line: 9449 F

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.