Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

The table failed to move down. The table failed to move down

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5164938 Tape Switch (208mm) used as IMS Tape Switch IN (rear) replaces 5127462-2 - used on GT2000/GT1700. See long description for details Part #: 5164938 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TAPE SWITH 4 ASSY Positioning GT FRU: Yes Status: Orderable Comments: added per to GT SPCN. source pole/planner code revised per to Global Table transfer Note: the little plastic screw for this part is U0087AA The Latest info is in this table. Switch Type Location Part Number (Latest) Switch Description in Block Diagram Tape Switch Part Number Description Switch Description in gesyslog Purpose Tape Switch TOP COVER FRONT ASSY 5127465-3 T-Sensor_Elev1 T-Sensor_Elev2 T-Sensor_Elev3 T-Sensor_Elev4 5133731 TAPE SWITCH 2 (185mm) Elevation touch switch1 (tape switches on underside of top front cover) Detection of clog in elevation down 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch1 (tape switches on underside of top front cover) Detection of clog in elevation down IMS LEFT SIDE COVER 5127463-2 T-Sensor_IMS3 5133730 TAPE SWITCH 1 (100mm) IMS tape switch OUT (front) Detection of pinching between IMS and column in elevation down T-Sensor_Elev10 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch4 (tape sw on underside of IMS side cover) Detection of clog in elevation down IMS REAR UPPER COVER 5127462-2 T-Sensor_IMS1 T-Sensor_IMS2 5164938 TAPE SWITCH 4 (208mm) IMS Tape Switch IN (rear) Detection of pinching between IMS and column in elevation up IMS RIGHT SIDE COVER 5127464-2 T-Sensor_IMS4 5133730 TAPE SWITCH 1 (100mm) IMS tape switch OUT (front) Detection of pinching between IMS and column in elevation down T-Sensor_Elev11 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch4 (tape sw on underside of IMS side cover) Detection of clog in elevation down Limit Switch GT REAR TOUCH SWITCH 5125112 T-Sensor_Elev7 T-Sensor_Elev8 T-Sensor_Elev9 N/A N/A Elevation touch switch3 (switch on underside of top rear cover) Detection of clog in elevation down TOUCH SENSOR R ASSY 5127469 T-Sensor_Elev5 N/A N/A Elevation touch switch2 (switches on underside of top side cover) Detection of clog in elevation down TOUCH SENSOR L ASSY 5127470 T-Sensor_Elev6 N/A N/A Elevation touch switch2 (switches on underside of top side cover) Detection of clog in elevation down
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues