GE Healthcare · CTLightSpeed200002712
LightSpeed 5.X (60kW)
Symptom
Scout images fail to reconstruct. See long description for details Scout images fail to reconstruct. Rebooting, cycling console power, etc. does not help.
Error message
Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies] See long description for full error. Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies]. The variable information gives the failing Group and Element and shows the failing string. Fri May 27 09:40:15 2005 Error: 200002712 Host: ct2 Process: imagecreate File: ic_mfm_main.c Line: 448 Image : 4243/1/1 Recon Sequence Id : 50 MFM error : Format failure Group Number : 40 Element : 254 Error status : -114 Error String : [ bad syntax for SH or LO or AT ]
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required