LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. See long description for pop up message picture. A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. After clicking OK Fast Cal completes. ATTENTION: The system detected a problem with the DAS/Detector sub-system. Image Quality will be compromised. Scanning will NOT be disabled, but it is critical that the user call GE SERVICE to rectify the problem. Press 'OK" to proceed at the risk of Image Artifacts
Error message
Error: 230007009 Exclusion Test - Row 1:32A, Channel 385, Module 25 Short to ground* PASSED CHECKS *. See LD for error string Sat Apr 23 12:49:15 2005 Error: 230023009 Host: erct Process: dailyPrepRx File: StateMachineEventNotify.c Line: 465 Tube temperature before Mylar Window Check is 461.80 degrees Celsius. Sat Apr 23 12:49:43 2005 Error: 230023009 Host: erct Process: dailyPrepRx File: StateMachineEventNotify.c Line: 465 Tube temperature before Bad Channel Map Update is 464.73 degrees Celsius. Sat Apr 23 12:50:30 2005 Error: 230007009 Host: erct Process: badchanmgr File: BadChannelMap.cxx Line: 659 Exclusion Test - Row 1:32A, Channel 385, Module 25 Short to ground* PASSED CHECKS *. Sat Apr 23 12:50:30 2005 Error: 230007009 Host: erct Process: badchanmgr File: BadChannelMap.cxx Line: 659 Exclusion Test - Row 28:5A, Channel 316, Module 20 Short to ground* PASSED CHECKS *. Sat Apr 23 12:50:30 2005 Error: 230007009 Host: erct Process: badchanmgr File: BadChannelMap.cxx Line: 659 Exclusion Test - Row 32:1A, Channel 678, Module 43 Short to ground* PASSED CHECKS *. Sat Apr 23 12:50:30 2005 Error: 230007009 Host: erct Process: badchanmgr File: BadChannelMap.cxx Line: 659 Exclusion Test
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5196401 FRDM for Center Modules for HDAS-64 see long description 5196401 HDAS FRDM Center Modules (Qty 57 used) Part #: 5196401 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: V-RES FIELD REPLACEABLE MODULE KIT - INNER SLOTS HALO CONVERTER CARD FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required